Size: 42X13X34 cm
Weight: 8.8 kg
HP 5335A Universal Counter
A high performance 200 MHz/2 ns universal counter
Built-in automatic rise time, duty cycle, pulse width,
slew rate and phase measurements
Advanced automatic triggering capabilities
HP-IB plus math and statistics functions standard
Specifications
Input Characteristics (channel A and B)
Range:
DC coupled, 0 to 100 MHz
AC 1 Mohm, 30 Hz to 100 MHz
AC 50 ohm, 200 kHz to 100 MHz
Sensitivity (00.001)
25 mV rms sinewave
75 mV peak-to-peak pulse at minimum pulse width at 5 ns
Dynamic Range (00.001, DC)
-5 V dc to +5 V dc
Trigger Level Range (00.001)
Auto Trigger OFF
Preset: set to 0 V dc NOMINAL
Adjustable: -5 v dc to +5 V dc
Auto Trigger ON
Preset: set to nominal 50% point of input signal
Adjustible: nominally between + and - peaks of input signal
Auto Trigger (00.001)
Range (50% duty cycle)
DC coupled, 30 Hz to 200 MHz
AC 1 Mohm, 30 Hz to 200 MHz
AC 50 ohm, 200 kHz to 200 MHz
Minimum Signal: 100 mV rms
Duty Cycle Range: 10% to 90%
Response Time: 3 second, typical
Impedance: 1 Mohm, shunted by < 35 pF or 50 ohm nominal
Frequency A
Range: 0 to 200 MHz, prescaled by 2
Period A
Range: 10 ns to 10 EXP 7 s
Time Interval A to B
Range: 0 ns to 10 EXP 7 s
Inverse Time Interval A to B
Range: 10 EXP -7 to 10 EXP 9 units/second
Rise and Fall Time A
Range: 20 ns to 10 ms transition with 50 Hz to 25 MHz
repetition rates (50% duty cycle)
Minimum Pulse Height: 500 mV peak-to-peak
Minimum Pulse Width: 20 ns
Duty Cycle Range: 20% to 80%
Pulse Width A
Range: 5 ns to 10 EXP 7 s
Trigger Point Range: 40% to 60% of pulse height
Duty Cycle A
Range: 1% to 99%, 0 to 100 MHz
Trigger Point Range: 40% to 60% of pulse height
Slew Rate A
Range: 50 V/s to 10 EXP 8 V/s slew rate with 50 Hz to 25 MHz
repetition rates (50% duty cycle). Minimum pulse height, width
and duty cycle range are same as rise and fall time A
Ratio A/B
Range: channel A, 0 to 200 MHz (prescale by 2)
channel B, 0 to 100 MHz
Totalize A
Range: 0 to 100 MHz
LSD Displayed: 1 count of input
HP-IB Output: at end of gate
Manual
Count Reset: via RESET key
Gated
Count Reset: automatic after measurement
Phase A rel B
Range: -180° to 360°, range hold off, or 0° to 360°, range hold on,
with signal repetition rates of 30 Hz to 1 MHz
Minimum Signal: 100 mV rms
LSD Displayed: 0.1°
Gate Time
Range: 100 ns to 10 EXP 7 s
Trigger Level
Range: x1, +5 to -5 V; x10, +50 to -50 V
Resolution: x1, 10 mV; x10, 100 mV
Time Base (with opt 010)
Frequency: 10 MHz
Short Term: < 1 x 10 EXP -10 rms for ls avarage
Warm-up: within 5 x 10 EXP -9 of final value in 20 min
Statistics
Sample Size: selectable between either n = 100 or n = 1,000 samples
Std. Dev: displays a standard deviation of selected sample size
Mean: displays mean estimate of selected sample size
Smooth: performs a weighted running avarage and truncates
unstable least significant digits from display
Math:
all measurement functions, with exception of GATE TIME, totalize in scale mode
and TRIG LVL, may be operated upon by math functions. Offset, Normalize, and
scale may be used independently or together as follows:
Number Value Range: ± 1 X 10 EXP -9 to ± 9 X 10 EXP 9
|